BEGIN:VCALENDAR
VERSION:2.0
PRODID:-//CERN//INDICO//EN
BEGIN:VEVENT
SUMMARY:Multichannel clinic dosimeter for the multiparameter direct contro
 l system of absorbed dose in areas of medical interest.
DTSTART;VALUE=DATE-TIME:20161014T121500Z
DTEND;VALUE=DATE-TIME:20161014T124500Z
DTSTAMP;VALUE=DATE-TIME:20260715T161443Z
UID:indico-contribution-554@cern.ch
DESCRIPTION:Speakers: Igor Abalakin (National Research Nuclear University 
 MEPhI)\, Valeriy Samosadny (National Research Nuclear University "MEPhI")\
 nLiterature research shows\, that plastic scintillator detectors\, includi
 ng their compact modifications\, provide most optimal characteristics for 
 the purpose of radiotherapy. These detectors can be used not only for in-v
 ivo on-line dosimetry\, but also in phantom tests during dosimetric planni
 ng systems and brachytherapy apparatus testing.\nMultichannel clinic dosim
 eter MKD-04 and several detectors using 1mm BSF-10 scintillator fiber have
  been designed. The detectors have 3.2 mm and 4 mm sensitive area diameter
 s. MKD-04 dosimeter provides dose rate measurement from 10 up to 5000 mGy/
 min.\nSome successful tests for the certification on the type of measuring
  instruments have been conducted. Clinical trials have been carried out on
  human phantoms using brachytherapy devices Agat-VT (Co-60)\, ROCUS-AM and
  Selectron (Ir-192).\n\nhttps://indico.particle.mephi.ru/event/4/contribut
 ions/554/
LOCATION:Milan Hotel Hall of the 2nd floor
URL:https://indico.particle.mephi.ru/event/4/contributions/554/
END:VEVENT
END:VCALENDAR