DESCRIPTION:Speakers: Dmitrii Kozlov ()\nRadiation hardness studies of the
  CMS detector's cathode strip chambers have been conducted at the GIF++ ga
 mma irradiation facility at CERN since 2016. During the studies\, ME1/1 an
 d ME2/1 chambers\, which differ in design and operate in CMS under high ba
 ckground rate\, were subjected to intense irradiation. The accumulated cha
 rge per unit length of the anode wire reached 1030 mC/cm for ME2/1 and 117
 0 mC/cm for ME1/1\, which is more than three times higher than the integra
 l dose expected for the entire period of HL-LHC operation. The characteris
 tics of the chambers were studied periodically in tests with SPS muon beam
 s. There are no signs of radiation-induced degradation of the chambers. Th
 e data obtained confirm the high radiation hardness of cathode strip chamb
 ers for operation in extreme conditions of HL-LHC.\n\nhttps://indico.parti