SUMMARY:The automated stand KALIF-8 for testing the characteristics of pho
 tomultiplier tubes
DTSTART;VALUE=DATE-TIME:20261027T162500Z
DTEND;VALUE=DATE-TIME:20261027T163000Z
DTSTAMP;VALUE=DATE-TIME:20261021T015422Z
UID:indico-contribution-6432@cern.ch
DESCRIPTION:Speakers: Georgy Akinin (National Research Nuclear University 
 MEPhI (Moscow Engineering Physics Institute))\nThis work is devoted to the
  restoration of the KALIF-8 automated test bench for evaluating the charac
 teristics of photomultiplier tubes of various sizes. During the work\, cab
 le lines were replaced\, the electronics were updated\, and the main compo
 nents of the measurement system were calibrated. The software for centrali
 zed control of the test bench was configured.\nDuring the experimental stu
 dies\, the main parameters of the reference photomultiplier tube were meas
 ured\, including gain\, linearity range\, timing and noise characteristics
 . To verify the reliability of the results\, a comparison was carried out 
 with measurements obtained using an independent test bench. The operabilit
 y and metrological suitability of the developed system for large-scale cal
 ibration of photomultiplier tubes prior to their installation in detectors